Initial Phantom Validation of Minute Roughness Measurement Using Phase Tracking for Arterial Wall Diagnosis Non-Invasively In Vivo

Författare:
Publiceringsår: 2011
Språk: Engelska
Sidor: 853-857
Publikation/Tidskrift/Serie: IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control
Volym: 58
Nummer: 4
Dokumenttyp: Artikel
Förlag: IEEE

Sammanfattning

To detect minute roughness, we utilized the phase change that occurs in a radio frequency echo from the rough surface of an object during its lateral motion. The new method was optimized and validated using saw-tooth-shaped silicone phantoms sized from 13 to 33 mu m; results were compared to those obtained using a confocal laser scanning microscope.