IFCC reference system for measurement of hemoglobin A(1c) in human blood and the National Standardization Schemes in the United States, Japan, and Sweden: A method-comparison study

Författare

  • W Hoelzel
  • C Weykamp
  • Jan-Olof Jeppsson
  • K Miedema
  • JR Barr
  • I Goodall
  • T Hoshino
  • WG John
  • U Kobold
  • R Little
  • A Mosca
  • P Mauri
  • R Paroni
  • F Susanto
  • I Takei
  • L Thienpont
  • M Umemoto
  • HM Wiedmeyer

Summary, in English

Background: The national programs for the harmonization of hemoglobin (Hb)A(1c) measurements in the US [National Glycohemoglobin Standardization Program (NGSP)], Japan [Japanese Diabetes Society (JDS)/Japanese Society of Clinical Chemistry (JSCC)], and Sweden are based on different designated comparison methods (DCMs). The future basis for international standardization will be the reference system developed by the IFCC Working Group on HbA(1c) Standardization. The aim of the present study was to determine the relationships between the IFCC Reference Method (RM) and the DCMs. Methods: Four method-comparison studies were performed in 2001-2003. In each study five to eight pooled blood samples were measured by 11 reference laboratories of the IFCC Network of Reference Laboratories, 9 Secondary Reference Laboratories of the NGSP, 3 reference laboratories of the JDS/JSCC program, and a Swedish reference laboratory. Regression equations were determined for the relationship between the IFCC RM and each of the DCMs. Results: Significant differences were observed between the HbA(1c) results of the IFCC RM and those of the DCMs. Significant differences were also demonstrated between the three DCMs. However, in all cases the relationship of the DCMs with the RM were linear. There were no statistically significant differences between the regression equations calculated for each of the four studies; therefore, the results could be combined. The relationship is described by the following regression equations: NGSP-HbA(1c) = 0.915(IFCC-HbA(1c)) + 2.15% (r(2) = 0.998); JDS/JSCC-HbA(1c) = 0.927(IFCC-HbA(1c)) + 1.73% (r(2) = 0.997); Swedish-HbA(1c) = 0.989(IFCC-HbA(1c)) + 0.88% (r(2) = 0.996). Conclusion: There is a firm and reproducible link between the IFCC RM and DCM HbA(1c) values. (C) 2004 American Association for Clinical Chemistry

Publiceringsår

2004

Språk

Engelska

Sidor

166-174

Publikation/Tidskrift/Serie

Clinical Chemistry

Volym

50

Issue

1

Dokumenttyp

Artikel i tidskrift

Förlag

American Association for Clinical Chemistry

Ämne

  • Biochemistry and Molecular Biology

Status

Published

Forskningsgrupp

  • Clinical Chemistry, Malmö

ISBN/ISSN/Övrigt

  • ISSN: 0009-9147