Supported liquid membrane extraction of urinary trans, trans-muconic acid, a biomarker for benzene exposure

Författare

Summary, in English

In this study, sample preparation of urinary trans, trans-muconic acid (tt-MA), a minor metabolite and biomarker of benzene, by means of Supported Liquid Membrane (SLM) extraction has been investigated. The sample extract was automatically transferred on-line to an ion exchange chromatographic column and the analytes were detected using a photodiode array detector. Automation of the system was accomplished using a robotic sample preparation processor integrated into the chromatographic system. Detection limits calculated from the baseline noise was in the sub-ppb range, and quantification of tt-MA is safe down to 10 ppb.

Publiceringsår

2002

Språk

Engelska

Sidor

351-355

Publikation/Tidskrift/Serie

Journal of Separation Science

Volym

25

Issue

5-6

Dokumenttyp

Artikel i tidskrift

Förlag

John Wiley & Sons Inc.

Ämne

  • Analytical Chemistry

Nyckelord

  • extraction
  • supported liquid membrane
  • benzene exposure
  • biomarker
  • muconic acid
  • urine analysis

Status

Published

ISBN/ISSN/Övrigt

  • ISSN: 1615-9314