Webbläsaren som du använder stöds inte av denna webbplats. Alla versioner av Internet Explorer stöds inte längre, av oss eller Microsoft (läs mer här: * https://www.microsoft.com/en-us/microsoft-365/windows/end-of-ie-support).

Var god och använd en modern webbläsare för att ta del av denna webbplats, som t.ex. nyaste versioner av Edge, Chrome, Firefox eller Safari osv.

Investigation of gamma-ray sensitivity of neutron detectors based on thin converter films

Författare

  • Anton Khaplanov
  • F. Piscitelli
  • J-C Buffet
  • J-F Clergeau
  • J. Correa
  • P. van Esch
  • M. Ferraton
  • B. Guerard
  • Richard Hall-Wilton

Summary, in English

Currently, many detector technologies for thermal neutron detection are in development in order to lower the demand for the rare He-3 gas. Gas detectors with solid thin film neutron converters readout by gas proportional counter method have been proposed as an appropriate choice for applications where large area coverage is necessary. In this paper, we investigate the probability for gamma-rays to generate a false count in a neutron measurement. Simulated results are compared to measurement with B-10 thin film prototypes and a He-3 detector. It is demonstrated that equal gamma-ray rejection to that of He-3 tubes is achieved with the new technology. The arguments and results presented here are also applicable to gas detectors with converters other than solid B-10 layers, such as Li-6 layers and (BF3)-B-10 gas.

Publiceringsår

2013

Språk

Engelska

Publikation/Tidskrift/Serie

Journal of Instrumentation

Volym

8

Dokumenttyp

Artikel i tidskrift

Förlag

IOP Publishing

Ämne

  • Accelerator Physics and Instrumentation

Nyckelord

  • Detector modelling and simulations I (interaction of radiation with
  • fast neutrons)
  • Gaseous detectors
  • thermal
  • Neutron detectors (cold
  • interaction of photons with matter
  • matter
  • interaction of hadrons with
  • etc)

Status

Published

ISBN/ISSN/Övrigt

  • ISSN: 1748-0221