Electroatriography - Time-frequency analysis of atrial, fibrillation from modified 12-lead ECG configurations for improved diagnosis and therapy

Författare:
Publiceringsår: 2007
Språk: Engelska
Sidor: 568-573
Publikation/Tidskrift/Serie: Medical Hypotheses
Volym: 68
Nummer: 3
Dokumenttyp: Artikel
Förlag: Churchill Livingstone

Sammanfattning

Current atrial fibrillation (AF) management guidelines suggest that initially a decision must be made to apply either a rate control or rhythm control strategy in the individual patient. However, patients' selection remains substantially empirical and the strategy initially chosen often proves unsuccessful and alternative therapies must be adopted. Thus, it seems desirable to develop and apply tests that quantify AF disease state and guide AF management. The overall hypothesis of this paper is that time-frequency analysis of AF from modified 12-lead ECG configurations will improve AF management beyond current diagnostic and therapeutic standards. In particular, we present a novel concept in which 12-lead ECG configurations are modified for time-frequency analysis of AF (electroatriography). While five electrodes (VR, VL, VF, V1, V2) are placed in the conventional position, the other four electrodes (V3, V4, V5, V6) are empirically repositioned anterior or posterior over the atria. By applying spatiotemporal QRST cancellation and time-frequency analysis to these recordings in 19 patients with persistent AF, fibrillatory rate dispersion among individual anterior (25 14 fibrillations per minute, fpm) and posterior leads (16 +/- 11 fpm) as well as individual anterior/posterior rate gradients ranging between -24 and +116 fpm could be identified. Consequently, the portrayed techniques may form the conceptual basis for individualized noninvasive characterization of AF. Initiation of further studies using the described techniques in different AF subsets, for comparisons with intracardiac recordings and outcome of different therapies, e.g. cardioversion, antiarrhythmic drug and ablation therapy may be stimulated. (c) 2006 Published by Elsevier Ltd.

Disputation

Nyckelord

  • Medicine and Health Sciences
  • Technology and Engineering

Övriga

Published
Yes
  • ISSN: 0306-9877

Box 117, 221 00 LUND
Telefon 046-222 00 00 (växel)
Telefax 046-222 47 20
lu [at] lu [dot] se

Fakturaadress: Box 188, 221 00 LUND
Organisationsnummer: 202100-3211
Om webbplatsen