Improved HPLC method for carbohydrate-deficient transferrin in serum

Författare

  • A Helander
  • A Husa
  • Jan-Olof Jeppsson

Summary, in English

Background: There is need for a reference method for transferrin glycoforms in serum to which routine immunologic methods for the alcohol marker carbohydrate-deficient transferrin (CDT) can be traceable. We describe an improved HPLC method for transferrin glycoforms. Methods: Transferrin was iron-saturated by mixing the serum with ferric nitrilotriacetic acid, and lipoproteins were precipitated with dextran sulfate and calcium chloride. Separation of glycoforms was performed on a SOURCE 15Q anion-exchange column using salt gradient elution. Quantification relied on selective absorbance of the iron-transferrin complex at 470 run. The relative amount of each glycoform was calculated as a percentage of the area under the curve, using baseline integration. Results: The HPLC system provided reproducible separation and quantification of the asialo-, monosialo-, disialo-, trisialo-, tetrasialo-, pentasialo-, and hexasialotransferrin glycoforms. Most importantly, disialo- and trisialotransferrin were almost baseline separated. The intra- and interassay CV for disialotransferrin were <5%. Serum and the pretreated samples were stable for at least 2 days at 22 or 4 degreesC. Sera from 132 healthy controls contained [mean (SD)] 1.16 (0.25)% disialotransferrin, 4.77 (1.36)% trisialotransferrin, 80.18 (2.01)% tetrasialotransferrin, and 13.88 (1.69)% pentasialo- + hexasialotransferrin. In some cases of a high (>6%) trisialotransferrin, monosialotransferrin was detected at <0.25%. Asialotransferrin was not detected in control sera, but was detected in 57% of chronic heavy drinkers and in 62% of sera with greater than or equal to2% disialotransferrin. Conclusions: The HPLC method fulfills the requirements of a preliminary reference method for CDT and should work for any combination of serum transferrin glycoforms. This method could also be useful for confirming positive CDT results by immunoassays in medico-legal cases. (C) 2003 American Association for Clinical Chemistry.

Publiceringsår

2003

Språk

Engelska

Sidor

1881-1890

Publikation/Tidskrift/Serie

Clinical Chemistry

Volym

49

Issue

11

Dokumenttyp

Artikel i tidskrift

Förlag

American Association for Clinical Chemistry

Ämne

  • Biochemistry and Molecular Biology

Status

Published

Forskningsgrupp

  • Clinical Chemistry, Malmö

ISBN/ISSN/Övrigt

  • ISSN: 0009-9147