Otoacoustic emissions and improved pass/fail separation using wavelet analysis and time windowing
Författare
Summary, in English
A new method is presented for the purpose of improving pass/fail separation during transient evoked otoacoustic emission (TEOAE) hearing screening. The method combines signal decomposition in scales using the discrete wavelet transform, non-linear denoising and scale-dependent time windowing. The cross-correlation coefficient between two subaveraged, processed TEOAE signals is used as a pass/fail criterion and assessed in relation to the pure-tone, mean hearing level. The performance is presented in terms of receiver operating characteristics for a database of 5214 individuals. The results show that the specificity improves from 68% to 83% at a sensitivity of 90% when compared with the conventional wave reproducibility parameter.
Publiceringsår
2001
Språk
Engelska
Sidor
134-139
Publikation/Tidskrift/Serie
Medical & Biological Engineering & Computing
Volym
39
Issue
1
Dokumenttyp
Artikel i tidskrift
Förlag
Springer
Ämne
- Other Medical Engineering
Status
Published
ISBN/ISSN/Övrigt
- ISSN: 0140-0118