Detection of hypotension during hemodialysis using the ECG

Författare:
Publiceringsår: 2004
Språk: Engelska
Sidor: 717-720
Publikation/Tidskrift/Serie: [Host publication title missing]
Volym: 31
Dokumenttyp: Konferensbidrag
Förlag: IEEE--Institute of Electrical and Electronics Engineers Inc.

Sammanfattning

There is today a lack of clinical techniques for detecting

acute hypotension during conventional hemodialysis treat

-ment, despite the fact that hypotension remains the - mosr common acute complication during hemodialysis.

Hypotension is often followed by nausea, vomiting, and

even fainting, not only strenuous for the patient but

requires considerable attention from the nursing staf. The

problem of detecting hypotension was studied by means

of a multimodal database. The database consists of

30 treatments in which each treatment includes several

simultaneously acquired signals. Acute .symptomatic

hypotension occurred in 2 of the 30 treatments. An

ECG-based method for detecting hypotension has been

developed. n e method involves information on heart

rate variability (HRV) and ectopic beat patterns. The

proposed method does not only detect the two cases of acute hypotension but also provides information of the patient's propensity to hypotension at an early stage of hemodialysis.

Nyckelord

  • Electrical Engineering, Electronic Engineering, Information Engineering

Övriga

Computers in Cardiology 2004
Published
  • Signal Processing-lup-obsolete
  • ISBN: 0-7803-8927-1

Box 117, 221 00 LUND
Telefon 046-222 00 00 (växel)
Telefax 046-222 47 20
lu [at] lu.se

Fakturaadress: Box 188, 221 00 LUND
Organisationsnummer: 202100-3211
Om webbplatsen