Calibration and Long-Term Stability of a PIXE Set-Up
Författare
Summary, in English
A general mass calibration procedure of a PIXE set-up is described. As an example, the results of the calibration of the PIXE set-up in Lund are given. For the calibration commercially available standards were used. The parameters of a physical model were adjusted to fit a calibration curve. To decrease errors due to inhomogeneities of the standard foils, the average from four different pieces of each foil was taken. The accuracy of the calibrated system in Lund is estimated to be better than 5% for individual elements and still less for ratios of adjacent elements.
For long-term quality control of the results, a special sample has been designed. This consists of a thick silver plate with a small copper spot in the centre. It is very sensitive to small changes in a PIXE set-up. By analysing this sample periodically, mistakes in the arrangement are promptly detected-a feature which is much appreciated in a mixed research, development and routine analysis laboratory. Also a measure of the long-term stability is obtained. For the Lund system, the long-term stability is 2.5% for a homogeneous sample and 4% for a small (inhomogeneous) sample.
For long-term quality control of the results, a special sample has been designed. This consists of a thick silver plate with a small copper spot in the centre. It is very sensitive to small changes in a PIXE set-up. By analysing this sample periodically, mistakes in the arrangement are promptly detected-a feature which is much appreciated in a mixed research, development and routine analysis laboratory. Also a measure of the long-term stability is obtained. For the Lund system, the long-term stability is 2.5% for a homogeneous sample and 4% for a small (inhomogeneous) sample.
Avdelning/ar
Publiceringsår
1981
Språk
Engelska
Sidor
81-88
Publikation/Tidskrift/Serie
Nuclear Instruments & Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors, and Associated Equipment
Volym
181
Issue
1-3
Dokumenttyp
Artikel i tidskrift
Förlag
Elsevier
Ämne
- Subatomic Physics
- Production Engineering, Human Work Science and Ergonomics
Nyckelord
- PIXE
- long-term stability
Status
Published
ISBN/ISSN/Övrigt
- ISSN: 0167-5087