A new automated method for analysis of gated-SPECT images based on a three-dimensional heart shaped model

Författare

Summary, in English

A new automated method for quantification of left ventricular function from gated-single photon emission computed tomography ( SPECT) images has been developed. The method for quantification of cardiac function (CAFU) is based on a heart shaped model and the active shape algorithm. The model contains statistical information of the variability of left ventricular shape. CAFU was adjusted based on the results from the analysis of five simulated gated-SPECT studies with well defined volumes of the left ventricle. The digital phantom NURBS-based Cardiac-Torso (NCAT) and the Monte-Carlo method SIMIND were used to simulate the studies. Finally CAFU was validated on ten rest studies from patients referred for routine stress/rest myocardial perfusion scintigraphy and compared with Cedar-Sinai quantitative gated-SPECT (QGS), a commercially available program for quantification of gated-SPECT images. The maximal differences between the CAFU estimations and the true left ventricular volumes of the digital phantoms were 11 ml for the end-diastolic volume (EDV), 3 ml for the end-systolic volume (ESV) and 3% for the ejection fraction (EF). The largest differences were seen in the smallest heart. In the patient group the EDV calculated using QGS and CAFU showed good agreement for large hearts and higher CAFU values compared with QGS for the smaller hearts. In the larger hearts, ESV was much larger for QGS than for CAFU both in the phantom and patient studies. In the smallest hearts there was good agreement between QGS and CAFU. The findings of this study indicate that our new automated method for quantification of gated-SPECT images can accurately measure left ventricular volumes and EF.

Publiceringsår

2005

Språk

Engelska

Sidor

234-240

Publikation/Tidskrift/Serie

Clinical Physiology and Functional Imaging

Volym

25

Issue

4

Dokumenttyp

Artikel i tidskrift

Förlag

John Wiley & Sons Inc.

Ämne

  • Physiology

Nyckelord

  • image processing
  • computer-assisted diagnosis
  • gated-SPECT
  • radionuclide imaging

Status

Published

Forskningsgrupp

  • Nuclear medicine, Malmö

ISBN/ISSN/Övrigt

  • ISSN: 1475-0961