Performance of two relatively new quantitative D-dimer assays (Innovance D-dimer and AxSYM D-dimer) for the exclusion of deep vein thrombosis.

Författare

Summary, in English

INTRODUCTION: D-dimer assays are now widely used as the first-line test in the diagnostic algorithm of suspected deep vein thrombosis (DVT). The aim of this study was to evaluate the performance of two relatively new quantitative D-Dimer assays (Innovance and AxSYM(R)) by comparison with a clinical gold standard. PATIENTS AND METHODS: 311 samples from outpatients with clinical suspicion of DVT, included in a prospective management study, was analysed (prevalence of DVT 23%). The diagnostic workup included estimation of pre-test probability, D-dimer determination, objective imaging as well as 3 month clinical follow up of negative patients. RESULTS: No significant differences were seen in sensitivity and negative predictive values between Innovance, AxSYM and the reference assays. The area under the ROC curve was slightly lower for the AxSYM assay and the correlation to the reference assays was only moderate (r<0.8) whereas the agreement with the Vidas assay was near excellent (kappa=0.8). The Innovance assay reached the highest AUC, showed a strong correlation with the reference assays (r>/=0.9) and a good agreement with the Vidas assay (kappa=0.76). In combination with a low pre-test probability score the Innovance assay reached a NPV of 100% (95% CI, 92-100) and the AxSYM assay 98% (95% CI, 87-100). CONCLUSION: The Innovance and AxSYM assays show an overall good and comparable performance for the exclusion of DVT when compared to the established assays. Our results for the AxSYM assay indicate that the optimal cut-off value needs to be further evaluated.

Publiceringsår

2009

Språk

Engelska

Sidor

701-705

Publikation/Tidskrift/Serie

Thrombosis Research

Volym

124

Dokumenttyp

Artikel i tidskrift

Förlag

Elsevier

Ämne

  • Cardiac and Cardiovascular Systems

Status

Published

Forskningsgrupp

  • Clinical Coagulation, Malmö
  • Clinical Chemistry, Malmö

ISBN/ISSN/Övrigt

  • ISSN: 1879-2472