First certified reference material for cystatin C in human serum ERM-DA471/IFCC.
Författare
Summary, in English
The IFCC Working Group for the Standardisation of Cystatin C (WG-SCC), in collaboration with the Institute for Reference Materials and Measurements (IRMM), announces the availability of the new certified reference material ERM-DA471/IFCC. The material was characterised using a pure protein primary reference preparation (PRP) as calibrant. The PRP was prepared from recombinant cystatin C, and its concentration measured using dry mass determination. The characterisation of ERM-DA471/IFCC was performed by particle enhanced immuno-nephelometry, particle enhanced immuno-turbidimetry, and enzyme amplified single radial immuno-diffusion. The certified cystatin C mass concentration in ERM-DA471/IFCC, if reconstituted according to the specified procedure, is 5.48 mg/L, the expanded uncertainty (k=2) being 0.15 mg/L.
Avdelning/ar
Publiceringsår
2010
Språk
Engelska
Sidor
1619-1621
Publikation/Tidskrift/Serie
Clinical Chemistry and Laboratory Medicine
Volym
48
Issue
11
Länkar
Dokumenttyp
Artikel i tidskrift
Förlag
De Gruyter
Ämne
- Clinical Laboratory Medicine
Status
Published
ISBN/ISSN/Övrigt
- ISSN: 1434-6621