Apparent Dead Space with the Anesthetic Conserving Device, AnaConDa®: A Clinical and Laboratory Investigation.
Författare
Summary, in English
The anesthetic conserving device (ACD) reduces consumption of volatile anesthetic drug by a conserving medium adsorbing exhaled drug during expiration and releasing it during inspiration. Elevated arterial CO2 tension (PaCO2) has been observed in patients using the ACD, despite tidal volume increase to compensate for larger apparatus dead space. In a test lung using room temperature dry gas, this was shown to be due to adsorption of CO2 in the ACD during expiration and release of CO2 during the following inspiration. The effect in the test lung was higher than in patients. We tested the hypothesis that a lesser dead space effect in patients is due to higher temperature and/or moisture attenuating rebreathing of CO2.
Avdelning/ar
Publiceringsår
2013
Språk
Engelska
Sidor
1319-1324
Publikation/Tidskrift/Serie
Anesthesia and Analgesia
Volym
117
Issue
6
Länkar
Dokumenttyp
Artikel i tidskrift
Förlag
Lippincott Williams & Wilkins
Ämne
- Anesthesiology and Intensive Care
Status
Published
Forskningsgrupp
- Integrative Health Research
ISBN/ISSN/Övrigt
- ISSN: 1526-7598