Compact medical fluorosensor for minimally invasive tissue characterization
Författare
Summary, in English
A compact fiber-optic point-measuring fluorosensor fully adapted to clinical studies is described. The system can use two excitation wavelengths, 337 and 405 nm, obtained from a nitrogen laser directly, or after dye laser conversion, respectively. The image intensifier used in the spectrometer can be gated with a variable time delay, allowing also time-resolved spectra to be extracted, with a time resolution of about 4 ns. Moreover, diffusely scattered white light can be spectrally recorded. The system is fully computer controlled enabling short recording times in clinical application, which are illustrated.
Avdelning/ar
Publiceringsår
2005
Språk
Engelska
Publikation/Tidskrift/Serie
Review of Scientific Instruments
Volym
76
Issue
3
Fulltext
- Available as PDF - 210 kB
- Download statistics
Dokumenttyp
Artikel i tidskrift
Förlag
American Institute of Physics (AIP)
Ämne
- Atom and Molecular Physics and Optics
Status
Published
ISBN/ISSN/Övrigt
- ISSN: 1089-7623