An off-axis STIM procedure for precise mass determination and imaging
Författare
Summary, in English
For the precise determination of the local mass density in thin samples analysed at the Lund Nuclear Microprobe, a modified version of off-axis STIM is described. An on-axis/off-axis geometry is used where a thin C-foil, placed behind the sample scatters protons into the particle detector. The advantage of this geometry compared to the off-axis case where particles are directly scattered from the sample into the particle detector is that the energy-loss spectrum acquired is less degraded. A spectral resolution comparable to on-axis STIM is achieved at a beam current high enough to allow for a simultaneous PIXE analysis (50-100 pA). (C) 2004 Elsevier B.V. All rights reserved.